Before performing any actions
, make sure the required calibration gas is applied
and flowing!
Supply all calibration gases with the same flow and pressure as the
sample gas (re-
commended: approx. 1 l/min), and utilizing
the correct gas fitting.
Ensure the warm-up time after switching on has elapsed! Warm-up time depends on
installed measuring system and configuration,
measurement specifications in
chapter 3!
Do NOT calibrate the TRACE MOISTURE sensor (tH
2
O)!
The sensor is completely calibrated with all calibration data stored in its flash memory
and does not require recalibration:
If the sensor is included into a calibration procedure, it might end up with a wrong ca
-
libration and unusable sensor. Therefore the analyzer's trace moisture measurement
channel is configured to be excluded from autocalibration procedures, by default calibra
-
ting all channels.
This exclusion is done by factory setup and cannot be changed.
For proper measurement results we recommend to exchange the sensor regularly after
12 months of operation. For instructions on how to exchange,
later in this chapter.
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do NOT
apply gas nor operate an internal pump before the warmup time has elapsed!
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!
Do NOT calibrate the TRACE OXYGEN sensor (tO
2
) without prior reading the
instructions!
Together with each sensor an installation manual is shipped, also giving comprehensive
calibration information.
Read these information
PRIOR intending to activate calibration procedures!
7.4.1 Preparing Calibrations
7.4.1 Preparing Calibrations