Emerson Process Management GmbH & Co. OHG
6-36
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Ch2
Procedure
CalProcMode:
AdvCal
Purge Time:
10 s
Max. Time:
120 s
Zero Ranges:
Together
Span Ranges
Separately
Zero Method:
Stability
Span Method:
Instant
Test Mode:
No
6.2.3.2.3
Setup Calibration Procedure
6.2.3 Setup Menu
If the system is setup accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
This menu line enables to select how the cur
-
rent channel is to be considered for calibration:
•
Selected measuring principles are permit-
ted to be calibrated considering special
conditions only, or are not permitted to be
calibrated at all.
•
Your process may require special hand-
ling of measuring channels with regard to
calibration.
Available options:
AdvCal
: This channel is permitted to be in
-
cluded into advanced calibration procedures.
Requires valves to be assigned!
SingleAuto
: This channel is permitted to be
calibrated by a single zero or span calibration
only.
Requires valves to be assigned!
Manual
: This channel is permitted to be ma
-
nually calibrated only.
Disabled
: This channel is not allowed to be
calibrated.
Notes!
Depending on your analyzer and selected
channel, not all options may be available.
E. g. for trace moisture sensors, calibration is
not permitted (
page 3-15 ).
AdvCal also enables SingleAuto and Ma-
nual calibrations.
SingleAuto also enables Manual calibrations!
Multi-channel unit:
In SELECT COMPONENT, select the chan-
nel to be configured.
Setup..
Calibration..
Procedure..
Note!
Marked lines are available only if “Valves” in
INSTALLED OPTIONS is set to a value other
than none.