Emerson Process Management GmbH & Co. OHG
6-66
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Signal:
Comp1
OutRange:
4-20 mA
Low Scale:
0.00
Max Scale:
100.00
AutoScale:
Yes
FailMode:
LOW - 10%
0/4 mA:
0.00
20 mA:
100.00
Hold:
No
6.2.3 Setup Menu
"
0/4 mA
" enables to finetune the analog out
-
put: Set "
Signal
" to
0 mA
and, while measu-
ring the output current, in this line adjust it to
the expected value.
Accepted range:
-10,000
...
+10,000
"
20 mA
" enables to finetune the analog out
-
put: Set "
Signal
" to
20 mA
and while measu-
ring the output current, in this line adjust it to
the expected value.
Accepted range:
-10,000
...
+10,000
"
Hold
" selects the output´s behaviour during
calibrations.
If set to
Yes
,
• the analog output is fixed to the last measu
-
red value;
•
concentration alarms, which may otherwise
be triggered by the concentrations of the
calibration gases, are supressed.
If set to
No
,
•
the analog output signal always corre-
sponds to the actual measured value during
calibration; this may trigger alarms if limits
are exceeded.
Note!
This behaviour may be undesireable if e.g.
the unit is connected to a data acquisition
system.