143
Reference Manual
00809-0100-4021, Rev JA
Configuration
April 2017
Configuration
4.5.4
Recommended actions for field diagnostics per NE107
4.5.5
Resource block diagnostics
Block errors
lists conditions reported in the BLOCK_ERR parameter.
Alarm type
Active event name
Recommended action text string
Maintenance
Required
Diagnostic Error
Device sensor diagnostic has been triggered.
Process Anomoly Detected
N/A
Out of Specification
Configuration Error
Re-write the sensor configuration.
Primary Value Degraded
Confirm the operating range of the applied sensor and/or
verify the sensor connection and device environment.
Calibration Error
Retrim the device.
Secondary Value Degraded
Verify the ambient temperature is within operating
limits.
Failed
Electronics Failure
Replace the device.
Asic Failure
Replace the device.
HW/SW Incompatible
Verify the hardware revision is compatible with the
software revision.
NV Memory Failure
Reset the device then download the device configuration.
Primary Value Failure
Verify the instrument process is within the sensor range
and/or confirm sensor configuration and wiring.
Secondary Value Failure
Verify sensor range and/or confirm sensor configuration
and wiring.
Function Check
Check
Transducer Block is under maintenance
Table 4-3. Resource Block BLOCK_ERR messages
Condition name
Description
Other
N/A
Device Needs Maintenance Now
N/A
Memory Failure
A memory failure has occurred in FLASH, RAM, or EEPROM memory.
Lost NV Data
Non-volatile data that is stored in non-volatile memory has been lost.
Device Needs Maintenance Now
N/A
Out of Service
The actual mode is out of service.
Summary of Contents for Rosemount 3144P
Page 2: ......
Page 10: ...8 Reference Manual 00809 0100 4021 Rev JA Title Page April 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 4021 Rev JA Introduction April 2017 Introduction ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev JA Installation April 2017 Installation ...
Page 235: ...225 Reference Manual 00809 0100 4021 Rev JA April 2017 ...