90
Reference Manual
00809-0100-4021, Rev JA
HART Commissioning
April 2017
HART Commissioning
Configure Rosemount X-well technology with AMS Device Manager
Perform the following procedure:
1. Right click on the device and select
Configure
.
2. In the menu tree, select
Manual Setup
.
3. Select the
Sensor
tab.
4. Select either
Rosemount X-well Process
or
Rosemount X-well Custom
.
5. Select desired configurations via
Basic Configuration
and select
Send
.
Figure 3-7. Manual Setup - Sensor Screen
3.9
Device output configuration
Device output configuration contains PV range values, alarm and saturation, HART output, and LCD
display options.;PV range values
Field Communicator
The PV URV and PV LRV commands, found in the PV Range Values menu screen, allow the user to set the
transmitter’s lower and upper range values using limits of expected readings. See
for unit and range setting limits. The range of expected readings is defined by the Lower Range Value
(LRV) and Upper Range Value (URV). The transmitter range values may be reset as often as necessary to
reflect changing process conditions. From the PV Range Values screen select
1 PV LRV
to change the
lower range value and
2 PV URV
to change the upper range value.
HART 5 Fast Keys
2, 2, 5, 5
HART 7 Fast Keys
2, 2, 5, 5
Summary of Contents for Rosemount 3144P
Page 2: ......
Page 10: ...8 Reference Manual 00809 0100 4021 Rev JA Title Page April 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 4021 Rev JA Introduction April 2017 Introduction ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev JA Installation April 2017 Installation ...
Page 235: ...225 Reference Manual 00809 0100 4021 Rev JA April 2017 ...