125
Reference Manual
00809-0100-4021, Rev JA
HART Commissioning
April 2017
HART Commissioning
5. The screen would display a summary screen of the coefficients
R0, A, B,
and
C
. Select
5 Set CVD
Coefficients
to set these coefficients.
6. When prompted, select which set of coefficients you would like to enter for that sensor. Select
between
1 R0, A, B, C,
and
2 R0, Alpha, Beta, Delta
.
7. When prompted, enter the desired values for each coefficient.
8. When you are done entering in these coefficients, another summary screen will appear. Review this
information, and when you are satisfied, select
OK
.
9. The method is complete, select
3 Exit
to exit the method if you are satisfied.