123
Reference Manual
00809-0100-4021, Rev JA
HART Commissioning
April 2017
HART Commissioning
6. This will reset any Min/Max values tracking this sensor and any Min/Max values tracking Differential,
Average, or First Good. Select
OK
.
7. It will now display the current CVD coefficients for the sensor (Alpha, beta, Delta, R0, A, B, C). Select
OK
.
8. Select which set of CVD coefficients you would like to enter for that sensor. Select between
1 R0, A, B,
C
, and
2 R0, Alpha, Beta, Delta
.
9. When prompted, enter in each constant and select
Enter
.
10. After you have completed this, it will display a summary screen with all the coefficient values needed
for the CVD equation. Review this information and select
OK
.