SE-1200 Series Electrocardiograph User Manual System Setup
- 88 -
printed on the different pieces of paper.
NOTE
:
Only when
Sampling Mode
is set to
Real-time Sample
,
Quickly
is available.
Record Sequence
Choose from:
Sequential
or
Synchronous
Select
Sequential
, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the
previous lead group.
Record Sequence
Select
Synchronous
, the lead group is printed one by one in a certain
sequence. All leads are printed with the same start time.
Gain
You can set the indicated height of 1mV ECG on the paper.
Choose from:
1.25mm/mV, 2.5mm/mV, 5mm/mV, 10mm/mV
,
20mm/mV
or
10/5mm/mV
.
10/5mm/mV
means that the gain of limb leads is set to
10mm/mV
, while
the gain of chest leads is set to
5mm/mV
.
AGC
(not available in
the VCG mode)
AGC
means auto gain control.
Choose from:
On
or
Off
Select
On
,
the gain can be automatically adjusted according to actual
signals.
Record Device
Choose from:
Thermal
,
HP1010/1510, HPM401,
HP1020/1020PLUS/1106
You should connect the corresponding USB printer
to the
electrocardiograph.
WARNING
If the printer used is not the type listed above, additional safety measures (such as
applying an isolation transformer to supply the medical system) should be taken when the
safety of the medical system has not been evaluated. If in doubt, consult our technical
service department or your local distributor.
CAUTION
It is forbidden to connect or disconnect a U disk or a USB printer during the transmission
course.
Record Device
NOTE
:
1. During the USB printing course
,
pressing the
PRINT/STOP
key
again cannot stop printing ECG reports.