SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 30 -
The electrocardiograph can be powered by either the mains supply or the battery.
To turn on the Electrocardiograph:
♦
When operating on AC power
Make sure that the mains supply meets the requirements (refer to A1.4 Power Supply
Specifications) before power-on, and then press
on the keyboard to turn on the
unit. The mains supply indicator ( ) is lit, and the logo will be displayed on the LCD
screen after self-test.
If the battery is weak when the mains supply is used, it will be recharged automatically
at the same time. Both the mains supply indicator ( ) and the battery recharging
indicator (
) will be lit.
♦
When operating on battery power
Press
on the keyboard to turn on the unit, and then the battery indicator (
) will
be lit and the battery symbol will be displayed. The logo will be displayed on the LCD
screen after self-test.
Because of the consumption during the storage and transport course, the battery capacity
may not be full. If the symbol
and the hint information
Battery Weak
are displayed,
which means the battery capacity is low, please recharge the battery first.
CAUTION
1. If the electrocardiograph is turned off because of low battery capacity or unexpected
power failure, the settings or the current ECG report may not be saved.
2. The electrocardiograph cannot print an ECG report when the battery is weak.
3. The use of electrocardiograph accessories (such as barcode reader) will deplete
battery power at a faster rate. The battery will require more frequent charging if these
accessories are used with the electrocardiograph.
To turn off the Electrocardiograph:
♦
When operating on AC power
Hold down the
key to display the hint
System is shutting down…
on the screen.
Then the device will be off a few seconds later. Remove the plug from the outlet.
♦
When operating on battery power
Hold down the
key to display the hint
System is shutting down…
on the screen.
Then the device will be off a few seconds later.