SE-1200 Series Electrocardiograph User Manual EMC Information
- 127 -
IEC/EN 61000-4-11 (60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
electrocardiograph
requires continued
operation during power
mains interruptions, it is
recommended that the
12-channel
electrocardiograph be
powered from an
uninterruptible power
supply or a battery.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration – electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The 12-channel electrocardiograph
is intended for use in the electromagnetic environment
specified below. The customer or the user of the 12-channel electrocardiograph
should assure
that it is used in such an environment.
Immunity
test
IEC/EN 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the
12-channel electrocardiograph,
including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC/EN
61000-4-6
3 V
rms
150 kHz to 80 MHz
3V
rms
P
d
2
.
1
=
Radiated RF
IEC/EN
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
P
d
2
.
1
=
80 MHz to 800 MHz
P
d
3
.
2
=
800 MHz to 2.5 GHz
Where
P
is the maximum output power