SE-1200 Series Electrocardiograph User Manual Printing ECG Reports
- 54 -
6.1.4 R-R Analysis Mode
The R-R analysis mode contributes to get the related indicators of HRV (Heart Rate Variability).
HRV analysis refers to the variability between durations of the two following heart rates, and it is
a predicted indicator to sudden death caused by acute myocardial infarction and vicious
arrhythmia. For healthy persons, RR Histogram presents open and multi-peak; for patient with
the low variability caused by cardiovascular diseases (such as myocardial infarction or heart
failure), RR Histogram presents a single peak.
Operation Method:
1.
To set
Mode Options
or
Rhythm Lead1
1)
Select
R-R
from the
Mode Option
list in the
Work Mode Setup
window.
2)
Select a lead from the
Rhythm Lead1
list in the
Lead Setup
window.
3)
Press
Enter
to confirm.
2.
Press the
PRINT/STOP
key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 180s, it begins to analyze and print an ECG
report.
3.
It will stop automatically after a complete R-R analysis report is printed. During the printing
course, pressing the
PRINT/STOP
key again can stop printing the ECG report.
NOTE
: In the R-R analysis mode, you cannot set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.