SE-1200 Series Electrocardiograph User Manual System Setup
- 98 -
Item
Description
Rhythm Lead1/2/3
Choose from:
І
,
П
,
Ш
,
aVR
,
aVL
,
aVF
,
V1
,
V2
,
V3
,
V4
,
V5
, or
V6
In the auto mode:
When
Record Style
is set to
3×4+1R
or
6×2+1R
, the rhythm lead
selected in the
Rhythm Lead1
list box will be printed in the ECG reports;
When
Record Style
is set to
3×4+3R
, 3 rhythm leads selected
respectively in the
Rhythm Lead1/2/3
list box will be printed in the ECG
reports.
In the rhythm mode:
When
Rhythm Style
is set to
Single Lead
, 60s wave of the rhythm lead
selected in the
Rhythm Lead1
list box will be printed in the ECG reports;
When
Rhythm Style
is set to
Three Leads
, 20s waves of three rhythm
leads selected respectively in the
Rhythm Lead1/2/3
list box will be
printed in the ECG reports.
In the R-R analysis mode:
The R-R analysis report of the rhythm lead selected in the
Rhythm Lead1
list box will be printed.
Lead Sequence Choose from:
Standard
or
Cabrera
Lead
Sequence
Lead group 1 Lead group 2
Lead group 3 Lead group 4
Standard
І
, II,
Ш
aVR, aVL, aVF V1, V2, V3
V4, V5, V6
Cabrera
aVL,
І
, -aVR
II, aVF,
Ш
V1, V2, V3
V4, V5, V6
Nehb
Choose from:
On
or
Off
.
Lead Sequence: I, II, III, ND, NA, NI
NOTE:
If you set
Nehb
to
On
, the working mode is fixed to be
manual.
Lead Off Hint
Choose from:
On
or
Off
.
When it is set to
On
and
lead off waves are detected in the presampled
waves, a lead off hint message will be displayed.
10.8 Display&Sound Setup
Select
Display&Sound
on the
System Setup
screen, and then press
Enter
to open the
Display&Sound Setup
window.