SE-1200 Series Electrocardiograph User Manual System Setup
- 93 -
Shift
+
Bksp
to refresh the patient information. The patient ID is 0~1999,
999, 999.
Select
Time
, press
Shift
+
Bksp
to refresh the patient information, and
then the patient ID can be automatically generated according to the time
when you press the
PRINT/STOP
key to print an ECG report. Entering
the patient ID manually is not supported.
ID Hint
Choose from:
On
or
Off
In the auto, rhythm or VCG (only configurable for SE-1200 Express)
mode, when
ID
is set to
Manual
and
ID Hint
is set to
On
, if you do not
input the patient ID before pressing the
PRINT/STOP
key, a hint will
pop up to remind you to input the patient ID.
Age
Choose from:
Age
,
D.O.B
or
Age Group
Select
Age
, you can enter the patient age manually in the
Patient
Information
window.
Select
D.O.B
, the
D.O.B
textbox appears and the
Age
textbox becomes
unavailable in the
Patient Information
window, you can enter the
birthday of the patient, and the system will calculate the patient age
automatically.
Select
Age Group
, the
Age Group
textbox appears in the
Patient
Information
window and the
0
key (or
Age Group
key) can be available.
For details, please refer to Section 2.2 “Keyboard and Keys”.
H/W Unit
Choose from:
cm/kg
or
inch/lb
BP Unit
Choose from:
mmHg
or
kPa
Prompt
Choose from:
Confirmed By
or
Unconfirmed
Select
Confirmed By,
the physician’s name is printed in the ECG reports
if it is input in the
Patient Information
window.
Select
Unconfirmed
,
Unconfirmed Report
is printed in the ECG reports.
PatInfo Refreshed
Choose from:
On
or
Off
Select
On
, the patient information will be refreshed after the ECG report
is printed out and all the leads are off.
Order Acquired
Choose from:
On
or
Off
Select
On
, the
Order
item will be displayed in the
Patient Information
window and you can acquire orders by clicking it.
User-defined
Input customized information such as
Address
, the information will be
displayed in the
Patient Information
window.