10.5 Patient Information Setup
Select
Patient Info
on the
System Setup
screen, and then press
Enter
to open the
Patient
Information Setup
window.
Figure 10-6 Patient Information Setup Window
Item
Description
Patient Options
Select
the item displayed in the
Patient Information
window.
Child Mode
is available only when Glasgow algorithm is used. In the
Child Mode, lead V3 is used to sample ECG signals of V4R.
The lead sequence of the Child Mode is: I, II, III, aVR, aVL, aVF, V4R,
V1, V2, V4, V5, V6.
NOTE:
1.
Pacemaker
appears in the
Patient Information
window after it
is selected in the
Patient Information Setup
window. Set
Pacemaker
to
Yes
in the
Patient Information
window, and the
Pacemaker
information will be displayed on the report printed
out.
2.
Pacemaker
is recommended to be set to
No
unless it is known
that the majority of the electrocardiograph usage will be on
patients with pacemakers.
3. Glasgow
algorithm
is an optional advanced function. To activate
it, please contact the local distributor.
ID
Choose from:
Auto, Time
or
Manual
Select
Manual
, the patient ID is within 30 ASCII characters.
Select
Auto
,
ID
can be automatically generated after you press