SE-1200 Series Electrocardiograph User Manual FAQ
- 108 -
Reload the paper in the paper tray, close the recorder casing firmly. If the problem still exists,
please contact the manufacturer or the local distributor for further disposal.
Q4: I want to print the hospital name in the report, but I can’t find the place to enter it, where is
it?
A4: Please open the
Other Setup
window, and move the cursor to the
Institution
textbox, and
then input the hospital name. The content you input in this textbox will be printed in the
report. For details, please refer to Section 10.12, “Other Setup”.
Q5: I pressed the
PRINT/STOP
key, but the ECG didn’t start printing, what’s wrong with it?
A5: The system will not respond to the
PRINT/STOP
key during the first 3s after you return to
the main screen. Therefore, you have to wait for a few seconds, and then you are able to start
the printing by pressing the
PRINT/STOP
key.
If you wait for a few seconds, but you still unable to start the printing by pressing the
PRINT/STOP
key, please check whether there is any error information displayed on the
screen.
If the hint
No Paper
or
Paper Error
is displayed on the screen, please deal with it according
to the above-mentioned measures.
If the hint
Transmitting…
is displayed on the screen, which means that the ECG is
transmitting the data to the PC, please wait a few seconds. You can start the printing after the
data is transmitted.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
Q6: I set the filter, speed and gain on the main screen1, but these settings were changed after
printing.
A6: The filter, speed and gain which are set on the main screen1 will not be saved, and they are
changed when you exit the main screen1 or after printing. If you want to save these settings,
please set them in the
Record Info Setup
window and the
Filter Setup
window.
理邦保密文件