SE-1200 Series Electrocardiograph User Manual Transmitting ECG Data
- 67 -
7.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF/DICOM Format
Through Ethernet Cable
NOTE:
The SCP/FDA-XML/DICOM function can be activated on the
Advanced Setup
screen. For details, please contact the manufacturer or the local distributor.
1.
Log into the FTP receiving software.
2.
Connect the electrocardiograph to the PC with an Ethernet cable recommended by the
manufacturer. If WIFI has been used, connect the PC and electrocardiograph to WIFI.
3.
Configure the
Transmission Setup
window.
1)
Press
Shift+F1
in the
Transmission Setup
window to open the
Basic Setup
window.
2)
Set
Auto Transmission
to
On
and
Transmission Mode
to
Wired
or
Wireless
.
3)
Set the
Server IP
item to the IP of the PC.
For details, please refer to Section 10.6.1 "Basic Setup".
4)
Set the
FTP User Name
,
FTP Password
and
FTP Path
items.
a)
The user name and the password you input in the
FTP User Name
and
FTP
Password
items must be available for FTP server.
b)
The path you input in the
FTP Path
item must be the subdirectory of the path you
input in the FTP receiving software.
NOTE:
For more information about FTP server, consult your Network
Administrator.
3.
Set
File Format
to DICOM/SCP/FDA-XML/PDF in the
File Setup
window.
4.
In the auto or rhythm mode, ECG data will be transmitted through the network automatically
after an ECG report is printed out.
7.2 Real-time Transmission to SE-1515
NOTE:
To use the real-time transmission function, the SE-1515 software of the
manufacturer must be installed in the PC.
1.
Double-click on the SE-1515 shortcut icon to start up the SE-1515 software..
1)
Enter the resting ECG sampling screen and click
Setup
. The System Setting window
will be display.
2)
Select the
Sampling Setting
tab, and set
Device Model
to
SE12
in the Sampling Box
setup panel.