SE-1200 Series Electrocardiograph User Manual System Setup
- 95 -
Item
Description
Device No.
If you input
0
in the
Device No.
textbox, after you save data to the U
disk or SD card, the data will be in the directory of
ECGDATA\ECG-0\Store\Examination Date
of the U disk or SD card;
after you export files from the electrocardiograph to the U disk or SD
card, the files will be in the directory of
ECGDATA\ECG-0\Export\Export Date and Time
of the U disk or SD
card.
Auto Transmission
Choose from:
On
or
Off
Select
On
, ECG data will be transmitted automatically after an ECG
report is printed out in the auto or rhythm mode; in the off mode, 10s
ECG data sampled before pressing the
PRINT/STOP
key can be
saved and transmitted, but cannot be printed.
Transmission Mode
(with WIFI
configured)
Choose from:
Wired, Wireless
, or
UART
Select
Wired
,
ECG data will be transmitted to the PC through the
wired network.
Select
UART
, ECG data will be transmitted to the PC through the
serial cable.
Select
Wireless
, ECG data will be transmitted to the PC through the
wireless network.
FTP Information
Enter data in the
FTP Path
,
FTP User Name
textboxes.
IP Addresses
Set Server IP, Local IP, Set Gateway, Set Subnet Mask
For details, please refer to Section 7.1.1 "Transmitting ECG Data in
DAT V1.03 Format".
NOTE:
If WIFI is enabled and
Auto Get IP
is selected in the
WIFI Setup
window, IP addresses can be acquired
automatically.
10.6.2 WIFI Setup (Optional)
Press
Shift+F2
to switch to the
WIFI
Setup
window.
理邦保密文件