SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
- 113 -
13.3.3 Disinfecting the Reusable Electrodes
1.
Wipe the suction bulbs of chest electrodes and the clamps of limb electrodes with a soft cloth
dampened with the disinfectant solution.
2.
Wipe off the disinfectant solution with a dry cloth after disinfection.
3.
Leave the suction bulbs and clamps to air dry for at least 30 minutes.
13.4 Care and Maintenance
CAUTION
Operate the cardiograph, charge the battery, and store the battery at a temperature of
40
℃
(104ºF) or lower. Exposure to higher temperature may reduce battery life, damage
the battery, and degrade overall cardiograph performance.
13.4.1 Recharge and Replacement of Battery
1)
Capacity Identification
The battery capacity can be identified according to the battery symbol in the top right corner
of the LCD screen.
->
->
->
Capacity is from full to empty.
2)
Recharge
SE-1200 series electrocardiograph is equipped with the recharge control circuit together with
the battery. When the unit is connected to the mains supply, the battery will be recharged
automatically. Then the battery recharging indicator (
) and the mains supply indicator
( ) will be lit at the same time. During the recharging course, the symbol
flashes in
the top right corner of the LCD screen. After the battery is fully recharged, the symbol
stops
flashing, and the battery recharging indicator (
) is black.
Because of the capacity consumption during the storage and transport course, the battery
capacity is not full when it is used for the first time. Battery recharge should be considered
before the first use.
NOTE:
The battery will automatically stop charging if you print an ECG report.
CAUTION
Repeated undercharging of the battery will damage the battery and reduce battery life.