SE-1200 Series Electrocardiograph User Manual Printing ECG Reports
- 61 -
(b)
The above figure (a) and (b) show an ECG report in the auto mode.
Template
is selected, and
Record Style
is set to
6×2+1R
.
The ECG report includes:
6×2+1R ECG waves
Current Date and Time
Patient Information:
Name, ID, Gender, Age
Measure Information:
HR
Heart Rate
P
P wave duration: the average P-wave duration from several selected
dominant beats;
PR
P-R interval: the average P-R interval from several selected dominant beats;
QRS
QRS complex duration: the average QRS complex duration from several
selected dominant beats;
QT/QTc
Q-T interval: the average Q-T interval from several selected dominant beats /
Normalized QT interval;
P/QRS/T Dominant direction of the average integrated ECG vectors;
RV5/SV1
The maximum of the amplitude of R or R’ wave of one selected dominant
理邦保密文件