SE-1200 Series Electrocardiograph User Manual Cleaning, Care and Maintenance
- 114 -
3)
Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please
contact the manufacturer or the local distributor for replacement.
WARNING
1. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and the battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
3. When the battery’s useful life is over, contact the manufacturer or the local distributor
for disposal or dispose of the battery according to local regulations.
4. Remove the battery from the electrocardiograph when the electrocardiograph isn’t
used for a long time.
5. If the battery is stored alone and not used for a long time, we recommend that the
battery be charged at least once every 6 months to prevent over-discharge.
CAUTION
If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.
13.4.2 Recorder Paper
NOTE
: Recorder paper provided by the manufacturer should be used. Other paper may
shorten the life of the thermal print head. The deteriorated print head may lead to
illegible ECG reports and block the advance of the paper.
Storage Requirements:
♦
Recorder paper should be stored in a dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
♦
Do not put the recorder paper under fluorescence for a long time.
♦
Make sure that there is no polyvinyl chloride or other chemicals in the storage
environment, which will lead to color change of the paper.
♦
Do not overlap the recorder paper for a long time, or else the ECG reports may
trans-print each other.