Guidance and manufacture’s declaration – electromagnetic immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The 12-channel electrocardiograph
is intended for use in the electromagnetic environment
specified below. The customer or the user of 12-channel electrocardiograph
should assure that
it is used in such an environment.
Immunity test
IEC/EN 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
±
2 kV for power
supply lines
±
2 kV for power
supply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC/EN 61000-4-5
±
1 kV line to line
±
2 kV line to
ground
±
1 kV line to line
±
2 kV line to
ground
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC/EN 61000-4-8
3A/m 3A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the 12-channel