SE-1200 Series Electrocardiograph User Manual System Setup
- 85 -
Rhythm Style
Select a style to print the ECG waves of the appointed rhythm lead.
Choose from:
Single Lead
or
Three Leads
Sampling Mode
(Only available in
the auto mode)
Choose from:
Pre-Sample
,
Real-time Sample
,
Triggered Sample
or
Periodic Sample
Select
Pre-Sample
, 10s ECG data sampled before pressing the
START/STOP
key will be printed out.
NOTE:
When
Sampling Mode
is set to
Pre-Sample
, if you press
the
PRINT/STOP
key before the electrocardiograph
samples for 10s, the recorder will not respond.
Sampling Mode
(Only available in
the auto mode)
Select
Real-time Sample
, 10s ECG data sampled after pressing the
PRINT/STOP
key will be printed out.
Select
Triggered Sample
, after pressing the
PRINT/STOP
key, if
Arrhythmia ECG data, including Asystole, Ventricular
Fibrillation/Ventricular Tachycardia, 5>PVCS>=3, Paired PVCS,
Bigeminy, Trigeminy, R ON T, single PVC and Missed Beat, is detected
during the learning course, the printing will be triggered automatically.
Select
Periodic Sample
, first you should set
Duration
and
Interval
.
NOTE:
ID and patient information will not be changed while carrying
out the periodic printing.
Duration
& Interval
In the auto mode, when
Sampling Mode
is set to
Periodic Sample
, if
Interval
is set to
2 min
,
Duration
is set to
24 min
, after pressing the
PRINT/STOP
key, the printing will be performed every two minutes and
come to 12 times.
Preview
Enable or disable the preview function in the auto or VCG (only
configurable for SE-1200 Express) mode.
Choose from:
On
or
Off
Auto Extend
Record
Choose from:
On
or
Off
Select
On
, if arrhythmia is detected in the auto mode, a hint will pop up to
ask you whether to print an extra rhythm report after the 12-lead ECG
report.
10.3 Filter Setup
Select
Filter
on the
System Setup
screen, and then press
Enter
to open the
Filter Setup
window.