Title:
DICTUM HEALTH IDM100 MEDICAL TABLET INSTRUCTIONS FOR USE
DCO#:
00271
Document#:
870-00001-01
Originating Author:
E Keate
Modifying Author:
R. Massie
Date Created:
9/04/2018
Date Modified:
05/28/2020
Revision:
AR
Page
144 of 175
2. Tap
NIBP Pump Pressure
on the SERVICE menu.
3. Tap
START
to close the Control Valve and the Dump Valve.
Ensure that the Hand Bulb Exhaust Valve is closed.
4. Increase the pressure to the normal NIBP range (approximately 280 mmHg) using the hand bulb.
5. Visually confirm that the measurement on the pressure device is within ± 3 % of the IDM100 Medical
Tablet pressure gauge measurement.