25.4 Pulse Oximeter Messages
Message
Cause
Corrective Action
WARNING:
SpO2 Not Functional.
A module error has
occurred.
Disconnect and reconnect the
cable/sensor. If the issue persists,
try a new cable/sensor pair.
If the error persists, contact Service.
WARNING:
Sensor Error Detected
The SpO2 sensor is faulty.
No SpO2
is connected.
The cable is faulty.
WARNING:
Communication Error
Detected
Contact Service.
WARNING:
Host Software Error
WARNING:
Pulse OX Control Board
Error
25.5 ECG Messages
Message
Cause
Corrective Action
WARNING:
ECG Cable Error.
Cable not connected.
Cable type unknown.
Cable is faulty.
Check connection for loose cable.
Use correct Dictum Health ECG
cable.
Replace faulty cable with a new
Dictum Health cable.
25.6 Spirometry Messages
Message
Cause
Corrective Action
WARNING
: Spirometer Board Revision
Not Supported
Contact Service.
WARNING
: EEPROM Checksum Error
WARNING
: Hardware Not Responding
WARNING
: Pressure Out of Supported
Range
WARNING
: Sensor Temperature out of
Supported Range
WARNING
: Timeout Error
WARNING:
Calibration Check failed.
Verify setup and repeat check. If check
fails again, discontinue use of
Spirometer and contact Service.
Discontinue use of Spirometer.
Follow the instructions in 21.5