Title:
DICTUM HEALTH IDM100 MEDICAL TABLET INSTRUCTIONS FOR USE
DCO#:
00271
Document#:
870-00001-01
Originating Author:
E Keate
Modifying Author:
R. Massie
Date Created:
9/04/2018
Date Modified:
05/28/2020
Revision:
AR
Page
173 of 175
Message
Cause
Corrective Action
WARNING:
Pneumatic Blockage.
Possible problem with the
cuff, hose or cuff
placement.
Check that the hose has no sharp
bends or is pinched.
Check that the patient is not lying
on the cuff.
Check that the cuff is in the correct
position.
WARNING:
BP reading terminated by
user.
Check the patient. Take another BP
reading.
WARNING:
Inflate Timeout, Air Leak or
Loose Cuff.
Possible problem with the
cuff, hose or cuff
placement.
Check that the hose is connected to
the system and the cuff.
Check that the cuff is properly
tightened.
Check that the cuff is in the correct
position.
Check that the correct size cuff is
being applied.
Check that the cuff is not leaking
air.
Check that the hose connections
are not damaged or loose.
WARNING:
Safety Timeout.
The patient may have
been moving too much.
Check the patient.
Check that the cuff is in the correct
position. Take another BP reading.
WARNING:
Cuff Overpressure.
Possible problem with the
cuff, hose or cuff
placement
Check that the correct size cuff is
being applied.
Check that the hose has no sharp
bends or is pinched.
Check that the cuff is in the correct
position.
Check that the patient is not lying
on the cuff.
WARNING:
Power supply out of range
or other hardware problem.
Service may be required.
Check that VPV and V
logic
are within
the voltage specifications. Check
the data/power input connection. If
applicable, check that sleep mode
operation is correct. Contact
Service.
WARNING:
Permission problem such
as safety link fitted or autozero out of
range.
Service may be required.
Check that LK1 jumper is not fitted.
Make sure proper command was
sent to module. Contact Service.
WARNING:
Transducer out of range.
Service may be required.
If possible, perform calibration.
Contact Service.
WARNING:
EEPROM calibration data
failure.
Service may be required.
If possible, perform calibration.
Contact Service.
WARNING:
MOSFET Failure.
Contact Service.