Title:
DICTUM HEALTH IDM100 MEDICAL TABLET INSTRUCTIONS FOR USE
DCO#:
00271
Document#:
870-00001-01
Originating Author:
E Keate
Modifying Author:
R. Massie
Date Created:
9/04/2018
Date Modified:
05/28/2020
Revision:
AR
Page
155 of 175
3. Tap
SERVICE
.
4. Tap
CALIBRATION
.
The Spirometer Calibration Check Setup screen appears.
5. Insert a new Spiraflow Sensor into the Spiragrip.
6. Connect the Spiragrip plug to the Spirometer port of the IDM100 Medical Tablet.
7. Push the mouthpiece of the Spiraflow Sensor all the way into the 3 L calibration syringe to ensure a
tight seal.
8. Pull the plunger out so that it is completely extended.
9. Place the syringe and Spiragrip on a clean, bare, level hard surface with 1 cubic meter of open space
minimum around the Spiraflow.