4. Increase the pressure to approximately 280 mmHg using the hand bulb.
To stop the test at any time, tap
STOP
.
5. Slowly increase the pressure from 280 mmHg to the overpressure point.
When the overpressure point is reached:
•
The valves open.
•
The pressure rapidly reduces to 0 mmHg.
•
The test ends.