If interpretation in ECG Settings is enabled, graphed results with measurement appear, as shown in the
example below. ECG calculated measurements return in addition when interpretation is enabled.
For more information, refer to Dictum Health
’s “IDM100 Diagnostic ECG Measurements for Clinicians”.
Caution:
If dashes, instead of numeric data, are returned, please note that the ECG
Calculated Measurements could be affected due to patient, or system, related issues.
(examples: when there are no ‘p’ or ‘t’ waves, if noise, artifacts, or out of range results, occur,
or if the ECG waveform cannot generate a normal beat). If dashes are returned instead of the
calculated measurements, check the patient’s ECG waveform. Check the patient and the
connections, and then repeat test.
Tap
SAVE
, and then disconnect the electrodes from the patient.
If interpretation is enabled in ECG Settings, the following parameters are returned as numeric data on the
ECG test results screen.
Parameter
Abbreviation
Parameter
Returned Data
Unit of Measure
Abbreviation
HR
Heart Rate
beats per minute
bpm
R-R
Interval between consecutive QRS
peaks
milliseconds
ms
P dur
P wave Duration
milliseconds
ms
PR int
Interval between P and R
milliseconds
ms