9.
Select
Previous menu - Fresh gas control
and perform step 7 and 8 but for N
2
O.
NOTE: If any of the pressure sensors are out of tolerance, they need to be recalibrated. For O
2
and air
pressure sensors, please refer to
“Part II”
section
“Electronic Ventilator - Calibration”
in this manual.
For the N
2
O sensor, refer to
“Part II”
section
“Fresh Gas Control Unit - Calibration”
.
NOTE: If any of the pressure gauges are out of tolerance, the need to be replaced.
Figure 40
The A-FGC1 and the A-EV1 diagnostic pages
10. Disconnect the test adapter gas inlet from the pipeline supplies gas (Air).
11. Wait one minute for the pressure in the system to stabilize.
NOTE: Make sure that the gas inlet assembly block on the A-WGU is in upward position when
performing the test of the non-return valves.
12. Check that the pressure drop on the test device does not exceed the following value during a
period of 2 minutes: < 0.07 x 100 kPa. (1 psi).
NOTE: If the check fails, check each gas line individually to find out which gas line is leaking and then
correct the failure. The total sum of pressure drop when testing the gaslines separately should not
exceed 0.2 x 100 kPa (3 psi) during a period of 2 minutes, which is equal to a pressure drop of 1 psi
when all three gas lines are tested at the same time.
13. Disconnect the high pressure test adapter.
14. Note the N
2
O pressure sensor value on the diagnostic page.
15. Check that pressure drop on N
2
O pressure sensor does not exceed the following value during a
2-minute period: < 0.2 x 100 kPa.
NOTE: If pressure drops more is this an indication of a leaking non-return valve in the WGU.
16. Select the previous menu and select Ventilator diagnostic page, and repeat steps 14 and 15 as
above but for O
2
and Air.
17. Connect the gas hoses O
2
, N
2
O and Air to their respective inlet connections on the WGU.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...