Agent flow calibration
This calibration calibrates the cassette flow sensor. The calculation is divided in two parts, one
second order and one linear calculation. The calibration factors for the safety limit calculation for the
agent proportional valve are also calibrated. The calibration factors are stored in EEPROM 25-29.
During this calibration the EEPROM 45 value is zeroed and it is calibrated at the first agent delivery
check after the calibration.
Equipment needed: Calibrated anesthesia monitor (Capnomac Ultima or
S/5 Anaesthesia with gas module can be used)
Filled enflurane or sevoflurane cassette
Settings:
ADU On
Pipeline gases connected
Manual mode
APL valve set to 1.5 cmH
2
O
No gas flows
Scavenging On
Figure 46
Agent flow calibration
1.
Push the
Setup
key. Select install and enter the distributor password. Select
Install - User
Config - Ambient pressure.
Select
Ambient pressure
and give external air pressure (see the
chart below). The first row shows the height above sea level (meters) and second row shows the
atmospheric pressure:
2.
Select
Service
and enter the calibration password. Check that all other calibrations are done
(see calibration tree in the beginning of this chapter). Then select
Calibrations - Vaporizer -
Agent flow.
Height (m)
0
200
400
600
800
1000 1200 1400 1600
Pressure (mbar)
1013 988
964
940
917
894
872
850
829
Height (m)
1800 2000 2200 2400 2600 2800 3000 3200 3400
Pressure (mbar)
809
789
769
750
732
714
696
679
662
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...