Troubleshooting A-FGC1
B-70
Document No. 8502854
N
2
O delivery test
Perce
ntage
of
test
Which valves are operated
What is being
tested
How test is done
Suggested actions if test fails
0 - 10
O
2
test valve is opened. The N
2
O
proportional valve is energized. The
N
2
O needle valve has to be manually
opened to over 6 l/min.
The N
2
O needle valve
and the N
2
O
proportional valve.
Looks for a flow of greater than 6 l/min.
through the sidegas flow transducer. If it can
not deliver enough flow, it remains in this
test.
Check O
2
and N
2
O pipeline pressure is
within specification.
O
2
test valve faulty.
11 - 40 The 02 test valve and the N
2
O
needle valve remain open. The N
2
O
proportional valve is closed.
This test verifies that
the N
2
O proportional
valve can shut off the
flow of N
2
O.
The sidegas flow transducer should indicate
a flow less than 60ml/min. If the test fails,
35%
is stored in the system check error
history.
Check flows with a Test device.
Check flows and N
2
O proportional valve
on diagnostic page.
2. Calibrate the A-FGC1.
41-50
The 02 test valve and the N
2
O
needle valve remain open. The N
2
O
proportional valve is reopened.
This test verifies that
the N
2
O proportional
valve has reopened.
The sidegas flow transducer should indicate
a flow of at least 5 l/min. If the test fails,
48%
is stored in the system check error
history.
Check the O
2
and side gas flows with a
Test device.
Calibrate the A-FGC1.
51-
100
The 02 test valve, the N
2
O
proportional valve and the N
2
O
needle valve remain open. The N
2
O/
Air selection solenoid is placed in
the Air position.
This test verifies that
selecting Air can shut
off the N
2
O flow.
The microprocessor in the Fresh Gas Control
unit de-energizes the N
2
O selection valve.
Looks for a flow through the sidegas
transducer. If there is still a flow, this
indicates that there is N
2
O flowing and it
has not switched to Air.
1 Check voltage to the N
2
O/Air solenoid
valve.
2. Replace the N
2
O/Air selection valve.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...