Calibration A-EV1
Document No. 8502855
C-35
CALIBRATION A-EV1
General requirements
•
The measuring equipment shall be accurate and calibrated
•
There are many types of flow and pressure test devices available that can be used to complete
this procedure. Datex-Ohmeda recommends no specific device. Specifications are provided
with this chapter. The flow and pressure test devices should have an accuracy of +/-2,5% or
better.
•
Datex-Ohmeda assumes that technical and biomedical professionals know how to use the
device available at his/her institution. Therefore, test device instructions are not included here.
•
If a check do not pass a test, verify and make sure that your test devices and its setup to the
measuring point are correctly done before starting to calibrate or changing parts.
•
Before flow calibration, ambient pressure on ADU shall be set on the average level of the region.
Also Manual and Auto leak check should be done.
•
The ADU is calibrated in ATP (ambient temperature and pressure).
•
Always makes sure to understand the function of the test devices before calibration.
•
Always remembers to not have any measuring tubes connected to the test device when you
Zeroing it.
•
The calibration procedure must follow the order as specified in the calibration tree below.
Entering calibration mode
Before performing any Ventilator calibration, make sure that the ventilator is in calibration mode:
1.
Enter the Service menu and
Calibration - Ventilator
.
2.
Check that the text
AUTO
or
MAN
has changed to
Calibrating
in the mode selection field
of the display.
3.
If not return to previous menu and select
Ventilator Diagnostic
menu.
4.
Enter pneumatic controls
CAUTION
Do not make any changes on valves
5.
Return to
Ventilator Calibrating
menu and check that the text in the mode selection field
has changed to
Calibrating
.
6.
Perform calibration.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...