![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 354](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507354.webp)
Datex-Ohmeda ADU
C-28
Document No. 8502855
Remounting the Electronic Ventilator to the ADU
Remounting is essentially reversing what was described in
“Dismounting the Electronic Ventilator unit
.
CAUTION
Make sure that the seals for the A-EV1 multi-connector and drive gas tube are
correctly seated before they are fastened.
After remounting, check that the primary driving gas setting is correct in the ventilator control unit
software. The setting can be checked on the
Setup - Service
menu.
If the setting is not correct, select
Setup - Install
menu using the distributor password. Switch the
primary gas to the correct one and check the setting on the service menu page.
Carry out the Functional Check to assure the functioning of the ventilator.
Replacing the EEPROM memory and A-EV1 software
In case of a CPU board (1) failure, move the EEPROM memory (2) or software (3) from the old CPU
board to a new CPU board, see Figure 22 on page C-28 and Figure 23 on page C-29. In this way there
is no need for a recalibration. If a new EEPROM is installed, a complete recalibration is necessary.
The EEPROM memory contains the module serial numbers and calibration data.
When ordering an EEPROM circuit as spare part, the serial number of the unit where it shall be
installed should be stated.
If the software or the EEPROM is replaced, carry out a Functional check.
To replace the EEPROM or software proceed as follows:
1.
Mains or gas supplies must not be connected to the unit.
2.
Remove the rear plastic cover of the ADU and the metal cover over the A-EV1.
3.
Replace the EEPROM (Chip socket D7 on circuit board) or the A-EV1 Software and refit the
covers. Make sure the software and EEprom will be installed correctly. Indication shall be
orientated as shown in Figure 23 on page C-29.
4.
Reconnect the pipeline gases and the mains supplies.
Figure 22
The A-EV1
1
2
3
4
5
6
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...