Datex-Ohmeda ADU
C-70
Document No. 8502855
Pilot pressure level check
The pilot pressure is measured at start-up, as it requires switching the MUX valve away from the
chamber. The pilot pressure is measured and if it is outside the specifications, a status message
‘Pilot press err’ (‘Low pilot press’) is triggered. The message appears on the
Service - Diagnostic
page’s status field. Accepted range is 250-310 cmH
2
O.
System checkout troubleshooting guide
If the Manual or Auto mode leak tests fails, the failure reason including the %failure number will be
stored in the
System Check - Check Log - Error History
page to give a better technical description of
the error that have been logged.
Manual mode leak test
error/
% no
Technical reason
Possible cause
Suggested action
Ppeak
= Pressure errors
5
Ppat >125 mmH
2
O
To high pressure. APL
valve not closed
properly.
- An extra pressure peak is detected during the test
(may have been caused accidentally while
handling the respiratory tubing or open a flow
control.)
-The pressure was higher than 20 cmH
2
O at the
start of the Manual Leak test. Open APL valve to
decrease the pressure before adjusting it to 80
cmH
2
O
56,66,76,
86
Pressure is rising in the
system.
98
Ppat (50 mmH2O) +
10 mmH
2
O < Ppat(80)
Pressure has slowly
increased.
Leak
= Leak errors
25
Ppat < 200 mmH
2
O and
V(leak) > 6000 ml/sec
To low pressure
because of leak.
- The manual APL valve is still open
- The respiratory tubing and/or its connections are
leaking
- The manual ventilation bag is leaking
- A gas monitor is connected to the circuit and is
sampling gas.
- Following Leak Tracing Chart for Auto Leak Test in
the
“Technical Reference Manual”.
26
Ppat < 200 mmH
2
O and
Ppat > 100 mmH
2
O
Patient pressure is not
enough after 60
seconds.
55,65,75,
85
Ppat < ¾ Pmax
Big leak in the system.
Pressure is below ¾ of
max pressure.
94
Ppat (50 mmH2O)
> Pat (80 mmH2O)
+ 50 mmH2O
A small leak during a
longer time.
100
Leak > 150mmH2O
Leakage to large.
Switch
= Auto/Man. switch errors
31,51,61,
71,81
Auto/Man. switch
changed during test.
- The position of the Auto/Man. selector has been
altered during the test. Repeat the test.
HiFlw
= Fresh gas flow detection errors
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...