![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 299](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507299.webp)
Troubleshooting A-FGC1
Document No. 8502854
B-73
20-29
The O
2
Test valve
is opened.
Scavenging valve
is closed and the
inlet valve is
opened.
Checks that the
scavenger valve
closes and that the
inlet valve is
opened.
For all agents except Desflurane, Is the counter pressure and
the cassette pressure compared, they should be within
30mmHg of each other.
The test for Desflurane is more rough and checks if the
difference between the cassette pressure and calculated vapor
pressure is larger than 500mmHg.
If the test fails will
28%
be stored in the system check error
history.
If failed at 28%
ALADIN cassette leaks or the connection
between A-FGC1 and cassette. Verify on A-
FGC1 diagnostic page.
Check cassette pressure sensor
calibration on the A-FGC1 diagnostic page
by verifying that
cassette pressure minus
install pressure = equal to counter
pressure (+30mmHg).
If not equal
calibrate Cassette pressure sensor, Agent
flow and repeat the test
Recalibrate Cassette pressure sensor
board.
Scavenging valve leaks; replace valve.
30-39
The inlet valve
and the O
2
test
valve remain
open.
Checks for a leak
in the cassette.
If the bypass flow is 4% less than the previous test (10-14%)
indicates this a leak in the cassette. Repeats the previous
cassette pressure test (20-29%). If the leak test fails will
33%
be stored in the system check error history. If the pressure test
fails will
35%
be stored in the system check error history.
If failed at 33%
Leakage in cassette or connection
between cassette and A-FGC1.
If failed at 35%
See suggestion at 28% 1-3
% of
agent
delivery
test
Which valves
are operated
What is being
tested
Description of test
Technical suggested actions
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...