![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 290](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507290.webp)
Troubleshooting A-FGC1
B-64
Document No. 8502854
Vaporizer
failure.
FGC safe state
after 2 minutes.
O
2
zero error
O
2
flow measurement zero
point drift 500 ADC. Note: Error
applies from Main SW version
6.1 and 7.1
O
2
flow measurement failure.
Replace the O
2
flow measurement zeroing valve.
Check the O
2
flow measurement unit.
x
Vaporizer
failure.
FGC safe state
after 2 minutes.
Sg zero error
Sidegas flow measurement
zero point drift 500 ADC. Note:
Error applies from Main SW
version 6.1 and 7.1
Sidegas flow measurement failure.
Replace the sidegas flow measurement unit Zeroing
valve.
Check the mixer flow measurement unit, sidegas.
x
Vapouriser
failure
FGC Mixer 0
unstab.
Zero value not stable during
zeroing
1) Leaking zeroing valve.
2) Faulty flow transducers.
1) Check the mixer zeroing valves and flow
transducers
x
Agent % not
delivered
Decr.flow.
FGC Insuffic.
agent
The measured ratio is
continuously smaller than the
set ratio and agent
proportional valve drive is at
maximum (>4000).
1) The required concentration is out of
reach at the current total flow and
cassette temperature.
2) Agent proportional valve is fully open
or stuck closed i.e. not cap-able of
increasing the agent flow due to an
agent channel occlusion or valve failure.
1) Decrease total flow or anaesthetic concentration
2) Check the agent proportional valve operation
and vaporizer maximum output.
x
N
2
O supply
pressure
lost
Measured N
2
O supply pressure
is below 2.4 bar
N
2
O supply pressure has decreased.
1) Increase the pressure.
Fresh gas
unit
failure
Safe state
Safe state: The AA and N
2
O
deliveries are interrupted. The
gas selection valve is switched
to Air position.
Major electronics or flow measurement
failure.
Message is always related to other
failure.
1) See related messages.
x
Message in
Normal
Screen
Message in
diagnostic
page
Error condition
Possible cause
Recommended
action
Err
log
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...