![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 101](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507101.webp)
Maintenance
Document No. 8502852
5-43
3.
Check that the fan works by cautiously inserting one hand into the cassette housing to feel the
airflow.
Turn the
On/Standby
switch to Standby position to leave the
Service
Menu and turn it on again.
Select
Bypass Check- Normal screen
and Confirm.
Step 17: Fresh gas outlet safety valve
Equipment needed:
Calibrated test device
Settings:
Pipeline supplies gases connected
Test Range:
0-150 cm H
2
O
Figure 49
Fresh Gas Outlet Safety Valve test setup
1.
Select the relevant low pressure range on the test device and connect a tube from the test
device inlet to the fresh gas outlet on the ADU.
2.
Adjust the total flow to 10 l/min. Check that the test device shows 80 - 100 cmH
2
O.
3.
Close the flows, disconnect the test device and reconnect the fresh gas hose to the fresh gas
outlet.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...