Technical description A-EV1
Document No. 8502855
C-15
Electronic Ventilator electronics
Sensors and electromechanical components are located inside the unit. The DC/DC power board,
CPU board and back-up processor, are concealed behind the unit.
The components and CPU board are connected to each other via a connection board and a cable. The
sensors have their own amplifier boards and the valves are connected directly to the connectors and
the board. A single cable connects the units to the CPU board.
All the electronic components are Electromagnetic Interference (EMI) protected inside a shielded
frame.
Data on driving parameters, patient circuit pressures, Auto/Man. switch position, delivered tidal
volume, used supply of driving gas and some alarms are sent to the ADU central electronics.
The control unit communicates with the ADU central electronics via a serial communication line.
The ventilator CPU board concealed behind the unit, consists of the power/current limiting circuits.
Components inside the unit is power limited to below 10 VA because of safety requirements. All
power limited driving signals are directed via flat cable from the ventilator CPU board to a ventilator
connection board.
DC/DC power board
The DC/DC power board, generates supply voltages for the CPU board and for the pressure sensor
boards; +5 V for digital circuits and
±
15 V for analogue circuits.
Figure 14
DC/DC power board
CPU board
The CPU controls the functions of the electronically controlled Ventilator. It controls the inspiratory,
expiratory and other valves, reads the pressure and flow signals from the transducers, sends the
information to the main CPU and receives commands from it through a serial interface.
The processor is a DS80C230, an Intel 51-derivative circuit, accompanied by external RAM for
runtime data, external EPROM for software storage and a serial EEPROM for calibration data.
The reset circuit includes power-up reset, supply voltage monitoring and watchdog circuit. The reset
signal always causes the outputs to go into off state.
Valve drivers
There are two kind of valve drivers, proportional and on/off -type.
In a proportional valve driver the current setting is analogous, so the current can be set anywhere
between zero and max. setting.
The on/off valve drivers are MOS-FET types where the load is connected between the output and the
ground.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...