Calibration A-FGC1
Document No. 8502854
B-49
Figure 47
Vaporizer agent flow calibration menu
3.
Insert the enflurane or sevoflurane cassette into the slot.
4.
Connect the monitor sampling port between the fresh gas hose and the fresh gas outlet.
5.
Insert the Aladin cassette ENF/AEV into the fresh gas control unit and flush the cassette
approximately 1 minutes by adjusting O
2
flow to 4 l/min. and Enflurane to 5.0%. After flushing,
turn agent off and close the flow.
6.
Start calibration by selecting
Start CAL.
7.
Adjust O
2
flow to the first calibration point. Let stabilize until the differences in anesthetic agent
concentration percentage on the monitor are maximum one decimal or until the agent valve
position on the delivery unit display has stayed unaltered for 5 seconds.
8.
Give reference percentage read from the external agent monitor by selecting
Give AA Conc
. and
turning
ComWheel
. Push
ComWheel
to enter the value.
9.
Repeat steps 7) and 8) for three more calibration points.
10. After entering the four calibration points, it is essential for successful calibration not to adjust
flow controls before the status ‘Calibration OK’ or ‘Calibration Failed’ is displayed
(approximately 1 minute). When the status ‘Calibration OK’ is displayed, save the calibration by
selecting
Save CAL
or abort the calibration by selecting
Abort CAL
. If the status is ‘Calibration
Failed’, abort the calibration by selecting
Abort CAL
.
11. Close flows and disconnect test equipment
12. Select
Previous Menu.
Perform the needed calibrations according to the calibration tree.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...