System description
Document No. 8502852
3-9
Fresh Gas Control Unit (A-FGC1)
The wall gas unit is connected to the gas inlets of the fresh gas control unit inside the gas distribution
block. The fresh gas control unit includes the flow controls and the vaporizer gas flow circuit.
Flow controls
The O
2
, N
2
O and Air flows are adjusted by means of mechanical needle valves.
The N
2
O flow may be adjusted up to 75% of total flow. This assures a minimum of 25% O
2
. Air may be
adjusted independently of O
2
flow.
Flow transducers electronically measure the O
2
, N
2
O and Air flows.
With the N
2
O/Air switch, the user may select either N
2
O or Air to be mixed with O
2
.
The mixed gas from the flow controls flows to the vaporizer gas flow circuit.
Electronically controlled vaporizer
The ADU is equipped with an electronically controlled vaporizer. The vaporizer gas flow circuitry and
agent concentration control system has been integrated within the delivery unit.
The gas flow from the flow controls is divided into bypass flow and liquid chamber flow. The gas
concentration is, through the control system, adjusted by changing the ratio between these two flows.
Aladin Cassette
The liquid chamber flow goes into the Aladin Cassette. This cassette is the only agent specific part of
the vaporizer. The agent cassette contains liquid agent, which is vaporized when the gas flow passes
through the cassette. The liquid chamber flow is then mixed with the bypass flow and conducted to
the fresh gas outlet.
There are keyed systems for filling the cassette with anesthetic agent. A level window indicates the
amount of liquid in the cassette. Changing the cassette, simply by pulling it out and inserting another
one, changes the agent.
Electronic Ventilator (A-EV1)
The ventilator consists of a control unit and a bellows unit.
The user may select the patient ventilation from two different modes of operation: Auto or Manual.
The selection is made by turning the Auto/Man. switch.
Auto mode
In Auto mode, driving pressure from the ventilator control unit, drives the bellows which ventilates the
patient via the patient circuit. It is possible to run the electronic ventilator in three ventilation modes:
VCV: Volume Controlled Ventilation
SIMV: Synchronized Intermittent Mandatory Ventilation
PCV: Pressure Controlled Ventilation
There are two sources of driving gas, primary (Air/O
2
) and secondary (O
2
/Air). If the pressure of the
primary gas is lower than 250 kPa (2.5 bar), the driving gas selecting unit automatically selects the
secondary gas for driving the ventilator.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...