![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 335](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507335.webp)
Technical description A-EV1
Document No. 8502855
C-9
The ventilator will go to safe state and send the message pressure compare error to main SW if the
differences between the Monitor and Circuit pressures are outside allowed limits.
•
Difference > 6.5 cmH
2
O for pressures over 55 cmH
2
O
•
Difference > 5.5 cmH
2
O for pressures over 25 cmH
2
O
•
Difference > 4.0 cmH
2
O otherwise.
Bellows chamber pressure sensor
The bottle pressure sensor is used for:
•
Cutting the inspiration if bellows reach the bottom
•
Measure pilot pressure during ventilator self test or manually by service diagnostics
•
Initiate the alarm Fill Bellows. This is done by comparing the patient pressure with the chamber
pressure value. In principal - if the chamber pressure is higher that the patient pressure, the
Inspiration interrupt and alarm are initiated. The table below describes the triggering limits for
the alarm Fill Bellows.
Gas consumption
In volume controlled mode the ventilator´s gas consumption is roughly the same as the patient
minute volume plus the waste fresh gas flow during expiration.
In PCV mode the ventilator will have higher gas consumption depending of set pressure, IE ratio and
flow.
As the compressed volume of the circuit is added to the volume delivered by the ventilator it should
be included in the gas consumption. E.g. when the compliance of circuit is 4 ml/cmH
2
O and the
circuit pressure is 15 cmH
2
O, the lost volume for each inspiration is 60 ml. If the ventilation rate is 12,
the extra gas consumption is 720 ml/min.
The PEEP function increases the gas consumption with the amount of the support flow. The required
support flow is about 10 l/min. As the consumption only increases during the expiration, it is
dependent of the I:E ratio setting.
Measuring sensors
The sensors are pressure sensors.
The chart below shows the sensors and their zeroing functions.
Patient peak pressure
Allowed pressure difference
(chamber pressure – patient pressure)
lower than 5.5 cmH
2
O
difference < 0.5 cmH
2
O
lower than 15 cmH
2
O
difference < (peak pressure – 5.0)
otherwise
difference < 10 cmH
2
O
Sensors and
measuring
Name on
diagnostic
page
Measuring
range
Zeroing
Patient circuit
Circuit
-40 - 100 cmH
2
O
zeroing valve 2 (zero phase 1 or 2)
Patient circuit (monitoring) Monitor
-40 - 100 cmH
2
O
zeroing valve 1 (zero phase 1 or 2)
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...