![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 190](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507190.webp)
Datex-Ohmeda ADU
A-38
Document No. 8502853
Removal of the WGU
Follow the steps below to remove the WGU.
Tools: 4 mm Allen wrench
1.
Disconnect the pipeline gases and close cylinder valves.
2.
Disconnect all external gas connections and reserve gas cylinders from the WGU.
3.
Turn the N
2
O/Air selector switch into N
2
O position and release N
2
O and Air pressures by first
opening the N
2
O flow fully and then opening O
2
flow to 1 l/min.
4.
Turn the N
2
O/Air selector switch to Air position and release Air pressure by opening the Air flow
control.
5.
Remove the rear plastic cover (10 screws).
6.
Remove the lower left side cover underneath the bellows block (? Allen screws).
Figure 33
Two multiconnectors mounted with four screws
7.
Remove the 10 remaining screws (plus washers) which hold the WGU to the frame.
8.
Gently pull out the WGU and put it on a bench.
9.
To reinstall after servicing, do the above steps in reversed order. Make sure the multiconnector
gaskets are in place before mounting them back the A-WGU.
10. Do a Functional Check. Replacement of Non-return Valves in the WGU
Replacing tubings or connectors
Always replace tubing that are damaged in anyway.
Always use D-O supplied tubing so correct type of material is maintained at its application.
Inspect the end of tubing before installing it to a connector, if the tube end are damage in any way can
it be cut straight with a racer sharp tool before connected. If the tube get to short so it gets tensed at
installation must it be replaced with a new tube.
When removing a tube from a quick connector the locking mechanism must first be pushed towards
the connector before the tube can be pulled out. See figure below.
Figure 34
Removing a tube from a quick connector
To install a tube to a quick connector, push the tube firmly into the bottom of the connector. Then it
will be locked (see figure below).
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...