Datex-Ohmeda ADU
B-56
Document No. 8502854
1.
Push the
Setup
key. Select
Service
and enter calibration password. Select
Calibration - Flow
controls - N
2
O & Air
. Connect the Test Device to the fresh gas outlet with a tube (Test Device set
to Air low flow range). Zero the Test Device before connecting the tube.
2.
Start calibration by selecting
Start CAL.
3.
Adjust air flow (read on the ADU display) to the first calibration point (1.0 l/min.). Let value
stabilize.
4.
Give reference flow (read on Test Device) by selecting Give Flow and turning
ComWheel
. Push
ComWheel
when the given air flow corresponds to the flow in Test Device.
5.
Repeat steps (4) and (5) for three more calibration points at 3.0, 7.0 and 11.0 l/min. (for 7.0
and 11.0 l/min. select air high flow range on Test Device and connect a larger tube). Acceptable
reference values are within
±
15% of the requested calibration points.
6.
When the status ‘Calibration OK’ is displayed, you can save the calibration by selecting
Save
CAL
or abort the calibration by selecting
Abort CAL
. If the status is ‘Calibration Failed’, abort
the calibration by selecting
Abort Cal
.
7.
Close Air flow control and turn N
2
O/Air Selector in position N
2
O.
8.
Setup Test Device to N
2
O flow range measurement, with the Test Device set for appropriate low
or high flow measuring range and repeat steps 4 to 7 with N
2
O.
NOTE: Maximum N
2
O flow 11 l/min. may not be attained even with the N
2
O flow control fully open.
This is not a fault proceed with the calibration as long you can attain 9.5l/min. If below 9.5l/min.
check first the pipeline inlet pressures is in specifications and secondly adjust the N
2
O regulator as
step A-C below.
A)
Remove the rear plastic cover (8 Allen screws), fold down the gas inlet part on the A-WGU (2
Allen screws) and open the rear cover of the fresh gas control unit (9 Pozidriv screws).
B)
Adjust the screw on the top of the N
2
O pressure regulator until the flow indicates 10l/min. on
the test device.
C)
Refit covers in reversed order after adjustment.
Figure 55
A-FGC1 Pressure regulators
9.
Select
Previous menu
. Perform the needed calibrations according to the Calibration Tree.
N
2
O
Air
O
2
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...