Technical description A-EV1
Document No. 8502855
C-1
TECHNICAL DESCRIPTION A-EV1
General
The ADU Electronic Ventilator module (A-EV1) is an electronically controlled, time-cycled flow
generator. The ventilator delivers tidal volumes compensated for the actual inspiratory fresh gas flow
and real measured breathing circuit compliance.
The A-EV1 includes:
•
The control unit
The ventilator control unit consists of hardware, electronics and software to control mechanical
ventilation and to monitor patient pressure.
•
The bellows unit
The ventilator bellows unit drives the breathing gas to the patient. Bellows unit also connects
pressure sensing lines to the patient circuit.
•
The user interface
The control parameters are set with the main display and keyboard.
Main ventilation modes
The ADU has two main ventilation modes; Auto or Manual. The mode selection is done by turning the
Auto/Man. switch, located on the bellows block, to Auto or Man. position.
Auto position turns the ventilator on and simultaneously routes the gases in the bellows to the
patient. The ventilator may work in three Sub-modes.
In Man. position the user routes the gases to the patient manually, by means of a reservoir bag.
Auto mode
In Auto mode, the ventilator unit ventilates the patient.
Three different ventilation sub-modes may be chosen by the user:
•
VCV = Volume Controlled (TV, MV)
•
SIMV = Synchronized Intermittent Mandatory Ventilation
•
PCV = Pressure Controlled Ventilation
See the User’s Reference Manual for detailed description of the modes.
The APL valve is not operating in Auto mode.
Manual mode
In Manual mode the operator ventilates the patient with a bag or the patient breathes spontaneously.
In Manual mode are the overflow valve and bellows not used.
The maximum circuit pressure is controlled by an Airway Pressure Limiting valve (APL). The valve is
located beside the Auto/Man. switch. When the pressure rises above the pre-set level, the APL valve
opens and gas bleeds to the scavenging system. When the patient breathes spontaneously, the
resistance of the APL valve is 1.5 cmH
2
O when the valve is set into minimum position.
Connections of the Electronic Ventilator module (A-EV1)
The connections of the module comprises pneumatic, mechanical and electrical connections.
•
The mechanical connection is between the frame and the module.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...