Datex-Ohmeda ADU
C-68
Document No. 8502855
In VCV mode with a set PEEP it may be observed that the pressure curve has an extra pressure rise
just before the expiration phase starts (see figure below). If the pressure rise is above 1cmH
2
O will
the problem be treated by:
1.
Check that the expiration valve unit membrane suffices side (the side against the seat) is dry
and without any grease. If the membranes feels little sticky can lime be rubbed in on the
membranes surfaces. After this should the membrane surfaces be bleed with air to remove all
loose particles from the lime. Reinstall the membranes to the expiration valve unit and
calibrated the PEEP following up with System Checkout.
2.
Replace the membranes in the expiration valve unit and calibrated the PEEP following up with
System Checkout.
Figure 41
Pressure curve
Description of ventilator self test
Electronics checks
When the ventilator starts working, the start-up code is executed. The code consists of checks and
initialisations.
Before any tests, there is a check if the ADU has been reset or properly switched on. If there has been
a reset, a program sets the unit in a safe state, immediately indicated by the alarm ‘Ventilator failure’.
If the ADU has been properly switched on, the operation of the most important electronic
components is checked.
Before any commands are accepted, the software version is checked between the main CPU and the
ventilator. Without the correct handshake, the system goes into safe state.
The electronics checks in order
1.
HW-watchdog check
2.
memory read and write check
3.
program check sum
4.
EEPROM value check
5.
Memory is zeroed and interrupts initialized.
6.
Valve exercise
The zeroing valves and both expiration valves has been implemented at start-up of the Ventilator to
increase the robustness of the self-test. The reason for this is that if an ADU has been out of use for a
prolonged period of time, there is a risk for sticking valves that can cause start-up problems of the
Ventilator during the self-test.
Exercising the zeroing valves will cause a rapid audible clicking sound during self-test. This
distinguishes version 7.6 from prior versions of software. Although it may seem unusual at first, it is
not a fault. The new self-test will take a total time of 70 seconds to execute.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...