Datex-Ohmeda ADU
C-80
Document No. 8502855
Leak tests for service purposes
These tests are for detecting leaks in specific areas of the ventilator and the bellows block. The test in
“Ventilator test (Test 1)” on page C-80, includes expiration and inspiration valves, flow measurement
unit, drive gas cylinder, bottle pressure line from the bellows block bottom plate to the chamber
pressure sensor and zeroing valve as well as PEEP support line and peep support valve in the same
way.
The test in “Test for circuit and monitor pressure sensor lines (Test 2)” on page C-80, involves the
pressure lines to the circuit and the monitor pressure sensors and their zeroing valves.
The purpose of the test in “Test for Bellows block and bottom plate leakage (Test 3)” on page C-81, is
to identify leakages in the bellows block and the bottom plate of the ADU for service purposes. The
following procedure is for testing the possible points of leakage in Auto ventilation mode.
Ventilator test (Test 1)
1.
The ADU in power On
2.
The gas inlets connected to the central outlets
3.
Enter Pneumatics Control on the Ventilator diagnostic page and Operate the expiration valves
(EXP. and PEEP on the diagnostic page). Close them both.
4.
Remove the bellows block.
5.
Connect a pressure manometer (cuff pump) to the driving gas hole in the bellows block bottom
plate with a suitable connector. See the figure below.
6.
Pressurize the volume with the cuff manometer up to 30 mmHg.
Figure 46
The channels in the Bellows Block bottom plate
7.
Wait for a minute. A pressure drop of 2 mmHg in 10 seconds at 30 mmHg pressure corresponds
approximately 10 ml/min. leakage. Pressure can also be observed from the output of bottle
pressure sensor on the diagnostics page of the ventilator.
Test for circuit and monitor pressure sensor lines (Test 2)
1.
Remove the bottom plate gasket.
2.
Connect the cuff manometer to the circuit and to the monitor pressure hole of the bellows
bottom plate. See the figure above.
3.
Pressurize the circuit lines to 30 mmHg.
NOTE: Do not apply too high pressure to the lines as it may damage the pressure sensors.
4.
A possible leakage can be detected by the pressure drop on the manometer or by the circuit
and monitor sensor values on the ventilator service menu.
Scavenging
Driving gas
PEEP support
bottle pressure
Circuit and monitor pressure
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...