![Datex-Ohmeda Anesthesia Delivery Unit S/5 Technical Reference Manual Download Page 437](http://html1.mh-extra.com/html/datex-ohmeda/anesthesia-delivery-unit-s-5/anesthesia-delivery-unit-s-5_technical-reference-manual_3764507437.webp)
Technical description A-ELEC
Document No. 8502856
D-7
Rectifier board
The function of the rectifier board is to rectify and filter the mains transformer output.
There are also comparators and switches that will turn off the output if the voltage is below 18 V or
above 36 V. The function of this feature is to protect the electronics from overvoltage and
undervoltage.
The output voltage of the board is VRAW
and it powers the ADU when on mains power.
Figure 6
Rectifier board
Battery charger board
The function of the battery charger board is to charge the batteries when the ADU is connected to the
mains and to supply power from the batteries to the VBUS if the mains is lost and the ADU is on.
There are two independent sets of batteries, (totally 4 batteries 12 V 2 Ah SLA-type) each having its
own charger circuit. The charger is of the constant current type with the output voltage limited to a
level suitable for standby use. As the charging voltage is temperature dependent, the temperatures of
the batteries are measured.
The charger circuits also include test loads for the batteries, so that the condition of the batteries can
be monitored. The CPU controls the test load, so that the battery voltage can be measured under
load. The charger circuits are disabled when the test load is applied.
The line comparator measures VRAW voltage and controls the battery switches. If VRAW drops below
20 V the batteries are connected to the VBUS and the ADU becomes battery powered. A signal from
the On/Standby switch is also controlling the function, so that the unit also can be switched on and
off when on battery power.
The CPU can also read a signal from the line comparator. If the line signal goes inactive a ”Mains lost”
alarm is generated.
A shutdown comparator measures the battery voltage. If it drops below 20 V a shutdown signal is
sent thus causing a shutdown alarm: ‘Battery empty, go manual‘. This also triggers the shutdown
timer which after a period of 30s, disconnects the batteries switches and the ADU is switched off.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...