Maintenance
Document No. 8502852
5-31
Functional check
Datex-Ohmeda recommends using the check list forms included with this manual.
Test devices
There are many types of flow and pressure test devices available that can be used to complete this
procedure. Datex-Ohmeda recommends no specific device. Specifications are provided with this
chapter. The flow and pressure test devices should have an accuracy of +/-2.5% or better.
Datex-Ohmeda assumes that technical and biomedical professionals know how to use the device
available at his/her institution. Therefore, test device instructions are not included in this manual.
NOTE: If a check does not pass a test, verify that your test devices and setup to the measuring point
are correct before starting to calibrate or changing parts.
A Functional Check must be carried out
•
After the first installation and after all reinstallations
•
Every six months
•
After planned maintenance
•
After maintenance procedures
•
After servicing the ADU
•
After storage
NOTE: If the ADU fails any part of the Functional Check, it must not be used in the operating theatre.
Equipment needed for a functional check
•
A calibrated test device to measure:
−
High/Low flow
−
High/Low pressure
•
Calibrated Gas Monitor (for example Capnomac Ultima, AM with Gas Module or 5330 Monitor)
•
Voltmeter
•
Aladin, Enflurane or Sevoflurane cassette
•
Electrical safety tester
•
Test lung 2 liters
•
High pressure test adapter, 8500002
•
Test Resistance when performing Tidal Volume test, 535335100
Step 0: Preparations
To prepare the monitor and test device for a Functional Check
1.
Turn on the gas monitor and the calibrated test device for a warm-up period.
2.
Calibrate the monitor with its calibration gas.
To prepare the ADU for a Functional Check
1.
Plug the power cord into a mains power supply outlet.
2.
Connect the ADU gas inlets to the hospital pipelines.
3.
Connect the ADU scavenging outlet to the hospital scavenging system.
4.
Connect the ADU vacuum outlet to the hospital vacuum system.
Summary of Contents for Anesthesia Delivery Unit S/5
Page 2: ......
Page 14: ......
Page 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Page 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Page 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Page 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Page 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Page 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Page 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Page 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Page 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Page 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Page 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Page 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Page 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Page 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Page 148: ......
Page 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Page 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Page 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Page 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Page 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Page 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Page 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Page 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Page 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Page 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Page 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Page 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Page 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Page 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Page 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Page 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...