11-2
ClosureRFG Radiofrequency Generator
Electrical Safety Tests 220-240 V
4. For the 25 A test, connect the ROD-L leads to the chassis ground lug, and to the
ground pin of the power cord.
Electrical Safety Test Limits 240 V
Parameter
Conditions
Passing
Input Voltage
Rated line + 10%
264 V
Line Cord Resistance
N/A
< 100 mΩ
25 A chassis-cord gnd
10 seconds
< 100 mΩ
Earth Leakage
Power Norm, GND OPEN
< 500 μA
Earth Leakage
Power Rev, GND OPEN
< 500 μA
Source Leakage, Applied Part
Power Normal, GND OPEN
< 10 μA
Source Leakage, Applied Part
Power Reversed, GND OPEN
< 10 μA
Iso Test, Applied Part
Power Normal
< 10 μA
Iso Test, Applied Part
Power Reversed
< 10 μA