![Covidien ClosureRFG Operation Manual Download Page 31](http://html1.mh-extra.com/html/covidien/closurerfg/closurerfg_operation-manual_2673930031.webp)
Chapter 3
ClosureRFG Radiofrequency Generator
3-1
Covidien ClosureFast Endovenous Radiofrequency
Catheter Treatment
Ready-to-Treat Screen
ClosureFast ready screen
Press Handle Button to Reset
ClosureFast
Total
00:00 m:s 0 RF Cycle(s)
1
2
3
4
5
6
Whenever a Covidien ClosureFast Endovenous radiofrequency catheter is connected, this
screen is present:
Catheter temperature:
Indicates current temperature of the catheter; it is not
intended to measure patient body temperature.
ཱ
Treatment Time:
Indicates, by counting down, the amount of time RF power has
been delivered during the current RF treatment cycle. For CLF catheters, the preset
treatment time is 20 seconds.
ི
Catheter Identifier:
“ClosureFast” is displayed.
ཱི
RF Ready Icons:
Indicates whether the ClosureRFG radiofrequency generator is ready
to start RF treatment. When dark or gray, RF power is disabled due to a limit violation.
When illuminated brighter, RF power is ready to be delivered to the catheter.